page_banner

News

How to Choose Clear and White PCR 8-strip Tubes in Your Lab?

Selecting the correct white or clear PCR 8-strip tubes can maximize sensitivity, ensure reproducibility, and ultimately, trust your results. To help you quickly assess the differences, here is a direct comparison of their key characteristics:

Feature White 8-strip Tubes Clear 8-strip Tubes
Light Interaction Reflective. Bounces fluorescent signal toward the detector. Transmissive. Allows signal to scatter and pass through the tube wall.
Data Sensitivity Higher. More signal captured often translates to lower Cq (Ct) values. Potentially Lower. Signal loss can lead to artificially higher Cq values.
Compatible Instruments Suitable for “Top-Reading” Optics instruments that detect fluorescence from above the tube.

Examples:

1. ABI StepOne / StepOne Plus / 7500 Fast / 7900 Fast

2. Bio-Rad CFX96 / CFX384 Touch Systems

3. Roche LightCycler® 480

4. Eppendorf Mastercycler® ep realplex

Suitable for “Bottom-Reading” or “Side-Reading” Optics instruments that detect fluorescence from bottom or side of the tube.

Examples:

1. ABI 7500 Fast  qPCR/ QuantStudio™ Systems

2. Roche LightCycler Nano

3. Shanghai Hongshi SLAN® Real-Time PCR Systems

4. Standard choice for conventional (end-point) PCR where no fluorescence detection is involved.

Looking for high-quality, certified PCR 8-strip tubes and caps? Browse Our Complete Range of White and Clear Options Here. 

Post time: Mar-05-2026